The EIS software techniques provide both status and error messages as they run. These messages are displayed in the lower left-hand corner of the runner window.
In all cases, an impedance data point was measured.
You are asked if you wish to retry the measurement, go to the next frequency, or terminate the entire experiment.
This failure is not fatal. The script resets the measurement settings and retries the measurement at the same frequency:
The voltage signal is significantly smaller than the value entered in Setup. The script will increase the FRA's signal output and try again. | |
The voltage signal is significantly bigger than the value entered in Setup. The script will decrease the FRA's signal output and try again. | |
The AC current signal is outside the FRA's range. The script will reset either the FRA sensitivity or the potentiostat's current range. | |
The DC current signal is outside the potentiostat's range or the FRA's common mode range. The script will reset either the FRA sensitivity or the potentiostat's current range. |