On the use of drift correction for electrochemical impedance spectroscopy measurements

By Mark E. Orazem & Burak Ulgut
Published in Electrochimica Acta 2023

Abstract

The effectiveness of drift correction is explored for measurement of electrochemical impedance spectra. Numerical simulations and experimental data for batteries and capacitors are used to show that drift correction cannot generate Kramers–Kronig consistent data for nonstationary systems. Application of drift correction improves, but does not fully recover, the original spectrum that is not corrupted by the drift. Drift correction improved the quality of spectra obtained for stationary systems under galvanostatic modulation, where the change from one frequency to another caused initial transients that corrupted the measurement.