Reduction of nonspecific protein adsorption on cantilever biosensors caused by transverse resonant mode vibration

By Johnson, Blake N. & Mutharasan, Raj
Published in Analyst The Royal Society of Chemistry 2014

Abstract

We examine if vibration of millimeter-sized cantilever sensors can release nonspecifically adsorbed proteins. Integrated electrochemical and mass-change measurement as well as fluorescence assays showed transverse surface vibration released nonspecifically bound proteins in samples prepared at 0.2-3.6 mg bovine serum albumin (BSA) per mL. Extent of release was directly related to magnitude of excitation voltage (Vex) applied to the self-actuating lead zirconate titanate (PZT) cantilever over three log units (0, 10 mV, 100 mV, and 1 V). Vibration-induced release was not instantaneous, but had an apparent first-order rate constant (kapp) which ranged from 0.02-0.1 min-1. Results suggest significant serum albumin protein release could be achieved using excitation voltages of 1 V in millimeter-sized cantilever sensors. Complementary experiments with thiolated DNA, which binds to surface gold ?111? sites with [similar] four times higher binding energy than BSA, showed negligible release under the same vibration magnitude. The results of the study suggest a direct correlation between surface-adsorbate binding energy and the effectiveness of vibration-induced release. We suggest that the release mechanism includes contributions from surface strain energy, body force, and acoustic streaming-associated hydrodynamic effects. The primary contribution of this study suggests that surface vibration of cantilever sensors may be useful in reducing nonspecific adsorption, especially for biosensing of analytes present in a complex background.

Read Article » Back